Abstract
Malaria rapid diagnostic tests (RDTs) have performed well in a variety of studies, but recent reports have described sensitivity for Plasmodium falciparum as significantly lower than that required for operational deployment. Exposure to high temperature has been suggested as an explanation. This study assessed the temperature stability of two different Plasmodium lactate dehydrogenase (pLDH)- and three histidine-rich protein 2 (HRP2)-detecting RDTs. One HRP2 test proved insufficiently sensitive for assessment. After incubation at 35, 45 and 60 degrees C, two RDTs detecting pLDH showed a substantial fall in percentage test line positivity over time, which was not seen with the remaining two HRP-2-based RDTs. For the particular products studied, variability was high, with the pLDH-based RDTs being less sensitive than HRP2-based RDTs against the sample of P. falciparum used and more susceptible to heat-induced damage, but the reasons for this are unclear. The performance of malaria RDTs can be adversely affected at the temperatures to which they will be exposed when transported to, and used in, the rural tropics.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Transactions of the Royal Society of Tropical Medicine and Hygiene
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.