Abstract
The GET electronics with the AsAd front end (based on AGET chips) was used to read out the auxiliary CeBr3 scintillation detectors of the SpecMAT active target. This article describes the influence of configurable parameters of the GET electronics on the recorded pulse-height γ-ray energy spectra. The lowest FWHM energy resolution is reported for sampling frequency in the range of 20–60 MS/s.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.