Abstract

An improved condensation nucleus counter (CNC) for use in the stratosphere is described. The University of Minnesota CNC (UMCNC) has a sequential saturator and condenser and uses n‐butyl alcohol as the working fluid. The use of a coaxial saturator flow, with aerosol in the center and filtered, alcohol‐laden air around it, speeds the response of this instrument and improves its stability as pressure changes. The counting efficiency has been studied as a function of particle size and pressure. The UMCNC provides an accurate measure of submicron aerosol concentration as long as the number distribution is not dominated by sub 0.02 μm diameter aerosol. The response of the UMCNC is compared with that of other stratospheric condensation nucleus counters, and the results of a (near) comparison with a balloon‐borne condensation nucleus counter are presented. The UMCNC has operated 14 times on a NASA U‐2 aircraft at altitudes from 8 to 21.5 km.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.