Abstract
Abstract. In this paper, the Photoacoustic Aerosol Absorption Spectrometer (PAAS-4λ) is introduced. PAAS-4λ was specifically developed for long-term monitoring tasks in (unattended) air quality stations. It uses four wavelengths coupled to a single acoustic resonator in a compact and robust set-up. The instrument has been thoroughly characterized and carefully calibrated in the laboratory using NO2/air mixtures and Nigrosin aerosol. It has an ultimate 1σ detection limit below 0.1 Mm−1, at a measurement precision and accuracy of 3 % and 10 %, respectively. In order to demonstrate the PAAS-4λ suitability for long-term monitoring tasks, the instrument is currently validated at the air quality monitoring station Pallas in Finland, about 140 km north of the Arctic circle. A total of 11 months of PAAS-4λ data from this deployment are presented and discussed in terms of instrument performance. Intercomparisons with the filter-based photometers of a continuous soot monitoring system (COSMOS), the Multi-Angle Absorption Photometer (MAAP), and Aethalometer (AE33) demonstrate the capabilities and value of PAAS-4λ, as well as for the validation of the widely used filter-based instruments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.