Abstract

The aerosol particle size distributions of solution-based pressurized metered dose inhalers containing 15%w/w ethanol and different quantities of nonvolatile component (NVC) (drug and glycerol) were evaluated at 25°C and 55°C, using a custom-built heating rig that preheated air prior to aerosolization. Particle size distributions were assessed using an Andersen cascade impactor and mass‐weighted cumulative aerodynamic diameter distributions were compared to a theoretical model that predicts the final size distribution, based on initial droplet size, vapor pressure of the formulation containing HFA 134a and percent NVC. In general, the mass median aerodynamic diameter was proportional to NVC1/3, with experimental particle size distributions following theoretical values. However, when comparing theoretical vs. experimental data over the range of mass-weighted cumulative aerodynamic diameter distributions between 10 and 90%, the 55°C experimental measurements more closely fitted the theoretical equation when...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.