Breath-alcohol analyzers used by law enforcement agencies to evaluate suspected driving-while-intoxicated (DWI) drivers are routinely calibrated with standard, dilute aqueous solutions of ethanol known as simulator solutions. The analyzers are deemed accurate if they generate results within an established margin of error consistent with ethanol concentrations equivalent to the actual concentrations of the simulator solutions. The fundamental flaw of this protocol is that it ignores the fact that a simulator solution is an ideal Henry’s law system, whereas a human test subject is not. Since breath-alcohol analysis is an application of Henry’s law, the level of accuracy ascribed to simulator-based calibrations cannot be applied to analyses involving humans because they are affected by physiological variables that can significantly impact the results of such analyses. This article addresses the importance of these variables, including blood:breath ratio variability, body temperature, and breathing pattern. M...
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.