Abstract

The point spread function (PSF) is used to characterize imaging systems. The PSF is usually not measured directly but rather the line spread function (LSF) is measured by scanning across the image of an input slit. One of the well known LSF-PSF conversion formulas is then applied.1 These formulas make the assumption that the length of the input-slit image is great compared to the PSF extent. This assumption is unfortunately unwarranted for one of the most important medical imaging devices: the x-ray image intensifier. The large extent image intensifier's PSF and the limited size of the intensifier's isoplanatic patches combine to make consideration of the finite length of the input slit important. For-mulas for calculating the PSF from a measurement of the finite-length line spread function (FLSF) have been developed for the case of a rotationally symmetric PSF.3 In this presentation we generalize the conversion formulas to cover non-symmetric PSF's.© (1988) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.