Abstract
Although D-dimer has gained widespread clinical use as a parameter for detection of in vivo fibrin formation, the issue of standardization of D-dimer assays remains to be resolved. The FACT study was performed to generate basic data for development of calibrators and standard preparations. A set of 86 samples, including plasma samples from patients with DIC, DVT. and other clinical conditions, serial dilutions of pooled plasma samples, and plasma samples containing fibrinogen- and fibrin derivatives, were distributed to 12 manufacturers of D-dimer assays. D-dimer assays differ concerning specificity for crosslinked fibrin, and preference for either high molecular weight fibrin complexes, or low molecular weight fibrin degradation products. Terminal plasmin digests of fibrin clots for calibration produce aberrant results in some assays, especially those with preference for high molecular weight crosslinked fibrin derivatives. The best conformity is achieved by the use of pooled plasma samples from patients with high levels of D-dimer antigen in plasma. In vitro preparations containing a comparable composition of fibrin derivatives to clinical plasma samples may also serve as reference material.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.