Abstract
Operation of the √f centroid detector with various doppler systems is discussed, and its design, implementation, and performance are described. It is determined that the detector can be employed in directional detection schemes to measure spectral moments and centroids with correct polarity and with no degradation in accuracy. Narrow- and wideband deterministic signal detection of both stationary and FM information is investigated, and it is found that the centroid detector can also serve as an FM demodulator with high AM rejection. A versatile √f detector was realized in discrete form and evaluated with deterministic and random signals, and its measured variance agreed closely with the theoretical values. The validity and application of the noise-reduction method was verified on deterministic and random doppler signals. Applications of the detector and the noise-reduction method to blood-flow and velocity measurements are demonstrated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.