The influence of the gas presence in calorimeter on the result of measuring of thermal capacity is considered. The feature of the case is the unavoidable change of the gas mass during measuring procedure. The correct expression for the value of amendment is obtained, and uncertainty is evaluated for the calorimeter КА-С4 that serves as the standard of the unit of heat capacity.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.