In this study, we propose a method of evaluating the spectral resolution of crossed-asymmetric Czerny-Turner spectrometers by comparing the impact of different slit functions on the optical transfer function, and different slit widths, through simulation and experiments. The results show that, the Gaussian function is suitable for narrower slit widths, such as 25 μm, and the rectangle function or convolution of slit-scattering function with rectangle function is suitable for wider slit widths, such as 50 μm, 150 μm, and 200 μm. The proposed method can provide guidance in the evaluation of spectral resolution in the preliminary optical design process of spectrometers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.