Abstract

To investigate the lubrication condition in knee prostheses under walking conditions, a knee joints simulator is used to simulate the tibial axis load and flexion-extension motion. The fluid film formation in knee prosthesis models is evaluated by measuring the contact electric resistance between a femoral stainless steel component and a tibial conductive elastomeric or polymeric component under constant applied voltage in a knee joint simulator. The effect of swinging motion and the variation of load on the fluid film formation is examined. It is shown that the fluid film formation in a knee joint with an elastomeric tibial component changes, depending on the walking phase, and the film thickness becomes minimum just prior to the toe off during walking motion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.