Abstract

This paper discusses the feasibility of a monitoring program for the quality assurance status of activity meters. We sent a questionnaire to clinical nuclear medicine departments of medical institutions, requesting information on their activity meters and quality assurance practices. On-site visits were conducted with exemption-level standard sources (Co-57, Cs-137 and Ba-133) for dose calibrators in nuclear medicine departments including physical inspection, accuracy and reproducibility. A method offering a quick check on the detection efficiency of the space dimension inside the activity meters was also introduced. For dose calibrator quality assurance, the daily checks had the highest implementation. However, annual checks and upon acceptance/after a repair check were reduced to 50% and 44%, respectively. The accuracy results of dose calibrators showed that all models exceeded the ±10% criteria with Co-57 and Cs-137 sources. The reproducibility results showed that some models exceeded the ±5% criteria with Co-57 and Cs-137 sources. The appropriate application of exemption-level standard sources considering the uncertainty that affects the measurement is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.