Abstract

We have examined the predictability of inspired isoflurane concentration during low-flow anaesthesia using a to-and-fro breathing system. Twenty one adult patients requiring mechanical ventilation of the lungs during surgery took part in this study. Using a fresh gas flow of 2 l.min-1, the ratio of inspired isoflurane concentration to isoflurane vaporizer setting was found to be approximately 4/5th after 5 min of anaesthesia. The ratio was maintained throughout the procedure, except for a few minutes following each change in vaporizer setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.