Abstract

Despite the excellent early performance of the Perceval sutureless valve, there have been recent reports of stent infolding, followed by paravalvular leak that was linked to valve oversizing. The purpose of this study was to review published reports by assessing photographic and radiological images. An in vitro study was also conducted with stent twisting that may occur during the collapse of the valve, termed 'stent entanglement', demonstrating consecutive successful valve collapse and implantation. This result has led us to hypothesize that infolding is due to a number of triggers, e.g. distortion of the stent overlooked during the final visual inspection of the implanted valve, rather than to excessive oversizing of the prosthesis as the sole cause, as repeatedly suggested in previous reports.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.