AbstractThis chapter deals with the signals produced by particles that are being absorbed in a calorimeter. The calorimeter response is defined as the average signal produced per unit energy deposited in this absorption process, for example in terms of picoCoulombs per GeV. Defined in this way, a linear calorimeter has a constant response. Typically, the response of the calorimeter depends on the type of particle absorbed in it. Also, most calorimeters are non-linear for hadronic shower detection. This is the essence of the so-called non-compensation problem, which has in practice major consequences for the performance of calorimeters. The origins of this problem, and its possible solutions are described. The roles of the sampling fraction, the sampling frequency, the signal integration time and the choice of the absorber and active materials are examined in detail. Important parameters, such as the e/mip and e/h values, are defined and methods to determine their value are described.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.