Abstract

To check the consistency of a (220)Rn measurement, performed by passive alpha track detector (ATD), the use of an accurate (220)Rn exposure calibration facility is mandatory. The ENEA Radon Service developed a new CR-39 ATD-Tn, coupled to the radon ATD-Rn and, to assess its sensitivity, had to design a small calibration facility, which has been recently modified to improve the spatial homogeneity exposure conditions inside the chamber, to get a continuous monitoring of the (220)Rn concentration and to reduce radon contamination. A better knowledge of the circuit response allowed selecting the best-operating conditions and how to regulate the thoron concentration. The active thoron monitor has been changed to serve as a continuous sampling and measuring device rather than a grab one; particular care has been devoted to assess the (216)Po losses on the device's inner surfaces and to evaluate the chamber transit time correction factor.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.