Abstract

Information is presented which, it is believed, should prove of value in eliminating, from electrocardiograph tracings, the effects of disturbances from electric lighting and power circuits. The direct cause of the disturbance was found to consist in certain defective electric motors or similar devices receiving current from the same lines as the building in which the electrocardiograph is situated. The solution of the problem lies in the location and repair or elimination of the defective machines.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.