The hydrogen maser for time keeping must have excellent performances with long term frequency stability, low frequency drift and high reliability for long term working. Based on the research experiences on hydrogen masers of Shanghai Observatory, the key technical improvements for time keeping hydrogen masers are described in this paper. The mainly electronic improvements include: analysis on the principle of the cavity auto-tuning technique and improving the modulation system, the development of phase-lock receiver. The primary test result reaches the expert performance as follows: the frequency stability reaches 1.5E-15/day, frequency drift rate is 8.6E-16/day, and temperature sensitivity is 8.1E-15/°.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.