A multi‐walled carbon nanotubes modified glassy carbon electrode (MWNTs/GCE) was fabricated, and the electrochemical behaviors of acetaminophen (ACOP) were investigated on the MWNTs/GCE. The results showed that MWNTs exhibited excellent electrocatalytic effects on the reaction of ACOP by accelerating the electron transfer rate. Cyclic voltammetry (CV) was used to explore the electrochemical redox mechanism of ACOP on the MWNTs/GCE and differential pulse voltammetry (DPV) was taken to determine ACOP in samples, respectively. The results showed that the oxidative peak currents were linear with the concentration of ACOP in the range of 4.0×10−7–1.5×10−4 mol l−1 with the detection limit 1.2×10−7 mol l−1. The MWNTs/GCE showed satisfactory stability, selectivity, and it can be used to quantify ACOP in effervescent dosage real samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.