Abstract

In general, externally powered prostheses do not provide proprioceptive feedback and thus require the user to rely on cognitively expensive visual feedback to effectively control the prosthesis. Applying the concept of extended physiological proprioception (EPP) to externally powered prostheses provides direct feedback to the user's proprioceptive system regarding the position, velocity, and forces applied to the prosthesis. However, electric elbows with EPP controllers developed at the Northwestern University Prosthetics Research Laboratory have exhibited unexplained "jerky" behavior in both clinical fittings and bench-top operation. In addition, the development of limit cycles, a specific type of constant-amplitude oscillation, had been observed in bench-top use of these elbows. Backlash and static friction within the EPP system were found to be primarily responsible for the development of limit cycles. Reducing static friction and backlash improved the system's performance. These results suggest that to most effectively implement EPP, prosthesis manufacturers should design prosthetic components that minimize static friction and backlash.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.