Abstract
The effects of potentiostat scan rate on the impedance, topography, and morphology of the Polyaniline (PANi) thin film has observed in this study. PANi has deposited on the Quartz Crystal Microbalance (QCM) surface with various scan rates, and changes in the impedance value have observed through an impedance analyzer test. Topography Measurement System (TMS) has observed the layer topography, while the layer morphology has observed using optical microscopy and Scanning Electron Microscope (SEM). The results have shown that the best sample has a scan rate of 10 mV/s, with a low impedance value indicating the layer has rigid. The variation in scan rate can affect the impedance value, but it is not significant and does not indicate a damping effect on QCM. A homogeneous layer is deposited at a low scan rate from topography and morphological observations, while a high scan rate results in an inhomogeneous layer.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: JPSE (Journal of Physical Science and Engineering)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.