Abstract

Variations in the shape and the power saturation of EPR spectra of L-alanine irradiated with photons, electrons, neutrons and protons are reported. It is shown that the ratio of the intensities of the satellite lines attributable to "spin flips" and the central line depend on the EPR microwave power, and it is proposed as a quantitative measure of the signal saturation effect. Dependence of this ratio on the microwave power is affected by the radiation quality and for doses in excess of 10 kGy by the absorbed dose. At high doses of low-LET radiation these changes are attributed to a high local density of free radicals, while for low doses of high-LET radiation these are due to changes induced in the crystal lattice. Consequently, the conventional peak-to-peak amplitude measurement of the EPR signal intensity is inaccurate when used for high doses and for comparison between radiations with different beam quality. The possibility to determine radiation quality from an EPR measurement is discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.