Abstract

Objective The study aims to investigate the effects of different adaptive statistical iterative reconstruction-V( ASiR-V) and convolution kernel parameters on stability of CT auto-segmentation which is based on deep learning. Method Twenty patients who have received pelvic radiotherapy were selected and different reconstruction parameters were used to establish CT images dataset. Then structures including three soft tissue organs (bladder, bowelbag, small intestine) and five bone organs (left and right femoral head, left and right femur, pelvic) were segmented automatically by deep learning neural network. Performance was evaluated by dice similarity coefficient( DSC) and Hausdorff distance, using filter back projection(FBP) as the reference. Results Auto-segmentation of deep learning is greatly affected by ASIR-V, but less affected by convolution kernel, especially in soft tissues. Conclusion The stability of auto-segmentation is affected by parameter selection of reconstruction algorithm. In practical application, it is necessary to find a balance between image quality and segmentation quality, or improve segmentation network to enhance the stability of auto-segmentation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.