Abstract

A systematic increase in computed tomography (CT) number of approximately 0.13 Hounsfield unit per scan (HU/scan) was observed when serial DeltaScan 2020 CT scans of a uniform water phantom were equally spaced at 0.5, 1.0, or 2.0 min and a shaped aluminum beam-hardening filter was employed. Much smaller drifts (less than 0.06 HU/scan) were observed with flat aluminum or shaped beryllium oxide filters. This machine drift, which was not associated with a rise in water phantom temperature and did not consistently correlate with estimated x-ray tube heat, could result in a significant overestimation of regional cerebral blood flow (rCBF) for a xenon/CT rCBF protocol involving 5-7 sequential scans obtained at 1-min interscan intervals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.