Abstract

The ICRU is considering revising the definition of ambient dose equivalent. This paper investigates the impacts of the proposed change on four designs of neutron survey instrument, the GNU, HSREM, LB6411 and Studsvik 2202D, in terms of their respective energy dependences of response and their performances in realistic workplace fields. In some circumstances the current designs of instrument still produce acceptable characteristics, but in general they may need to be re-optimized to better match the requirements of the new operational quantity; to that end, a simple retrofit solution for the GNU is demonstrated. The performance criteria against which instruments are judged may also need to be revised to reflect the proposed change.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.