Background Despite endoscope reprocessing, residual droplets remain in gastrointestinal endoscope working channels. Inadequate drying of gastrointestinal endoscope working channels may promote microbial reproduction and biofilm formation, increasing the risk of infection in patients. This review was designed to provide the current status of gastrointestinal endoscope drying, emphasize the importance of gastrointestinal endoscope drying, and evaluate the effectiveness of different drying methods of gastrointestinal endoscope in reducing residual droplets and microbial growth risk. Methods A systematic review was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) reporting checklist. The PubMed, Web of Science, Medline, EMBASE, EBSCO, CNKI, CQVIP, and Wanfang Data databases were searched from 2010 to 2020 to identify eligible articles focused on methods of gastrointestinal endoscope drying and the status of endoscope drying. The following key points were analyzed: type of intervention, amount of residual droplets, major microbial types, and effectiveness of biofilm intervention. JBI quality assessment tool was used to determine bias risk for inclusion in the article. Results This review included twelve articles. Two of the articles reported lack of drying of gastrointestinal endoscopes while the other ten reported residual droplets, microbial growth, and biofilm formation after different methods of drying. Four articles reported 0 to 4.55 residual droplets; four articles reported that the main microbial types were cocci and bacilli, most commonly Staphylococcus, Escherichia coli, Bacillus maltophilia, and Pseudomonas aeruginosa; and two reported that drying could effectively reduce biofilm regeneration. The type of intervention is as follows: automatic endoscopy reprocessor (AER), manual compressed air drying, and the Dri-Scope Aid for automatic drying and drying cabinet. Conclusions While endoscope reprocessing may not always be effective, an automatic endoscope reprocessor plus the Dri-Scope Aid with automatic drying over 10 min or storage in a drying cabinet for 72 h may be preferred.

Highlights

  • Gastrointestinal endoscopy plays an important role in the prevention of digestive diseases [1]

  • This review showed that the extent of endoscopic drying was not sufficient in clinical practice, as recommended in existing guidelines for gastrointestinal endoscope reprocessing; the guidelines or statements for endoscope reprocessing around the world are inconsistent, and no clear and uniform requirements have been made on the duration and method of drying

  • Gastrointestinal endoscopes still had residual droplets, microbial growth, and biofilm formation after reprocessing and drying, but we found that automatic endoscope reprocessor (AER) plus Dri-Scope Aid automatic drying was obviously better than automatic endoscopy reprocessor (AER) plus manual drying and the use of drying cabinet once again reduced the residue of droplets and the growth of microorganisms, improving the quality of gastrointestinal endoscope reprocessing

Read more

Summary

Introduction

Gastrointestinal endoscopy plays an important role in the prevention of digestive diseases [1]. Residual droplets remain in gastrointestinal endoscope working channels. Inadequate drying of gastrointestinal endoscope working channels may promote microbial reproduction and biofilm formation, increasing the risk of infection in patients. This review was designed to provide the current status of gastrointestinal endoscope drying, emphasize the importance of gastrointestinal endoscope drying, and evaluate the effectiveness of different drying methods of gastrointestinal endoscope in reducing residual droplets and microbial growth risk. Two of the articles reported lack of drying of gastrointestinal endoscopes while the other ten reported residual droplets, microbial growth, and biofilm formation after different methods of drying. While endoscope reprocessing may not always be effective, an automatic endoscope reprocessor plus the Dri-Scope Aid with automatic drying over 10 min or storage in a drying cabinet for 72 h may be preferred

Methods
Results
Discussion
Conclusion