Abstract

IntroductionThe aim of this study was to evaluate the stability of urine collected in preservative tubes for chemistry strip analyses and particle counting to determine whether the transport of urine samples with all of their constituents is possible.Materials and methods275 pathologic urine specimens were included. Each urine sample was evaluated after 4, 8, 12, 24, and 48 hours of storage in BD Vacutainer® Plus Urinalysis Preservative (BD UAP) tubes and compared with refrigeration at 4 °C. All analyses were peformed on H-800 and FUS-200 automatic modular urine analyzers (Dirui Industry, Changchun, China). The kappa coefficients (κ), false positive (FP) and false negative (FN) rates were evaluated. κ > 0.8 was accepted as good agreement.ResultsHaemoglobin (Hb), leucocyte esterase (LE), and protein (Pro) analyses should be performed within 4 hours, whereas glucose (Glc) was stable until the end of 48 hours in both storage conditions. Nitrite (Nit) was well preserved in BD UAP tubes for 24 hours but was stable only up to 8 hours at 4 °C. Bilirubin (Bil) had very high FN rates even at 4 hours in both conditions. The particle counting showed high FN rates for white blood cells (WBC) and red blood cells (RBC), whereas squamous epithelial cells (EC) were stable up to 8 hours in both conditions.ConclusionsPreanalytical requirements for both urine chemical strip analyses and particle counting in a unique sample were not met in either condition. Thus, the transfer of urine samples for centralization of urinalysis is not yet feasible.

Highlights

  • The aim of this study was to evaluate the stability of urine collected in preservative tubes for chemistry strip analyses and particle counting to determine whether the transport of urine samples with all of their constituents is possible

  • Haemoglobin (Hb), leucocyte esterase (LE), and protein (Pro) analyses should be performed within 4 hours, whereas glucose (Glc) was stable until the end of 48 hours in both storage conditions

  • The particle counting showed high false negative (FN) rates for white blood cells (WBC) and red blood cells (RBC), whereas squamous epithelial cells (EC) were stable up to 8 hours in both conditions. Preanalytical requirements for both urine chemical strip analyses and particle counting in a unique sample were not met in either condition

Read more

Summary

Introduction

The aim of this study was to evaluate the stability of urine collected in preservative tubes for chemistry strip analyses and particle counting to determine whether the transport of urine samples with all of their constituents is possible. Automated urine analysers, including chemical strip readers and visual microscopic testing, have increased the efficiency of routine laboratories [2,3]. Despite improvements in the performance of analytic systems, the pre-analytical phase of modern urinalysis remains obscure [1]. In both chemical strip analyses and particle counting, the sample must be kept stable until the analysis. Solutions to the problem of stability of urine analytes will make the centralization of urine testing possible

Objectives
Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.