Abstract

The effect of volume on the liquid scintillation counting performance of 14C-samples has been investigated. A decrease in counting efficiency was observed for samples with volumes below about 6 ml and those above about 18 ml when unquenched samples were assayed. Two quench-correction methods, sample channels ratio and external standard channels ratio, and three different liquid scintillation counters, were used in an investigation to determine the magnitude of the error in predicting counting efficiencies when small volume samples (2 ml) with different levels of quenching were assayed. The 2 ml samples exhibited slightly greater standard deviations of the difference between predicted and determined counting efficiencies than did 15 ml samples. Nevertheless, the magnitude of the errors indicate that if the sample channels ratio method of quench correction is employed, 2 ml samples may be counted in conventional counting vials with little loss in counting precision.