Abstract

Acoustic radiation force (ARF) elastography has recently become available. The previous animal studies have revealed lung injuries induced by diagnostic ultrasound, but the effects on the lung resulting from exposure to ultrasound with ARF are unknown. This study aimed to assess the risk of lung injury associated with ultrasound with ARF. A focused 2.5-MHz transducer that emits ultrasound with ARF was used. A rabbit was anesthetized, and the transducer was placed in the right subcostal region. Exposure settings of mechanical index (MI) 0.80, pulse duration 10ms, pulse repetition time 5s, and exposure time 150s were applied. One red spot (7×6mm) was observed on the surface of the right lung corresponding to the area of exposure. Alveolar hemorrhage was observed microscopically. This lesion was visible across a range of 20-170μm in depth from the pleural surface. The first example of lung hemorrhage induced by ultrasound with ARF was observed in this study. This observation suggests the possibility of lung injury in humans when ARF elastography is applied with the transducer directed toward the lung. Further studies are needed to determine the safety of this modality.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.