Abstract
We tested the effect of reducing the ultrasound output energy on obstetric measurements made from two-dimensional (2-D) grayscale images. We measured three fetal structures (femur, umbilical vein and lateral cerebral ventricle) in 113 pregnant women, commencing with an output energy that provided a thermal index for bone (TIB) of ≤1.0, and then repeated the measurements at TIB ≤ 0.5 and then TIB ≤ 0.1. There was a trend toward shorter femur measurements and larger umbilical vein diameter measurements with decreasing output energy. However, since these effects (which did not exceed 0.03 mm) were very small compared with the inter- and intraobserver variations, they were considered to be of no clinical significance. Furthermore, the residuals did not increase when the output energy decreased, indicating that the precision of the measurements was unaltered. We conclude that using a reduced output energy setting corresponding to a TIB of 0.1 does not compromise ultrasound biometry findings.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.