Abstract
Rigid body mechanics with computer data acquisition and analysis techniques were used to determine the three-dimensional motions of any point on the tibial component of a total knee arthroplasty. Three stem configurations were compared: (1) no stem; (2) short stem (40 mm); and (3) long stem (75 mm). In addition, three loading conditions were analyzed for each stem configuration: (1) central loading; (2) posterior loading; and (3) medial loading. The longer stem implants were associated with increased micromotion, especially under eccentric loading. Cemented implants seemed to have more stable fixation, compared with noncemented implants. It was thought that the increased motion was secondary to a toggling of the implant under load, secondary to uneven medullary cortical contact. Overall, the results indicated that short and long stems do not enhance initial fixation with cemented or cementless implantation in routine knee arthroplasty.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.