Abstract

This study aimed to evaluate whether dual-energy computed tomography can reduce metal artifacts and improve detection of pulmonary nodules. Twelve simulated nodules were randomly placed inside a chest phantom with a pacemaker. Then, dual-energy computed tomography was performed, and 5 virtual monochromatic images at 40, 50, 65, 100, and 140 keV were reconstructed with 5- and 0.625-mm slice thicknesses. Two independent observers assessed the metal artifact (3-point scale from 1, none, to 3, severe) and detection of the nodule (5-point scale from 1, definitely absent, to 5, definitely present). Statistical analysis was performed with a P value of less than 0.01 (0.05/5). With both slice thicknesses, the metallic artifact increased at 40 or 50 keV and decreased at 100 or 140 keV relative to that at 65 keV (P < 0.01). The nodule detection score was not significantly different between each kiloelectron volt level with the 0.625-mm slice thickness; however, the score was significantly worse at 40 keV compared to 65 keV (P < 0.01) with the 5-mm slice thickness. High monochromatic energy images can reduce metal artifacts without a change in nodule detection score. Low monochromatic energy images increase metal artifacts and worsen nodule detection in thick slices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.