Abstract

Surface functional electrical stimulation results in stimulation of cutaneous receptors and discomfort. The degree of non-uniformity of current distribution in the cutaneous layers in the vicinity of the electrode may influence the sensation experienced. This paper describes the effects on sensation of a thin, high impedance electrode designed to reduce the non-uniformity of current distribution. Sensation associated with stimulation via a self-adhesive electrode with much higher impedance than conventional electrodes was compared with a low impedance electrode in a single-blinded, crossover study. The high impedance electrode does not alter either the current at which sensation is first registered. However, at higher currents, the high impedance electrode allows 9% more current to be passed for an equivalent sensation to that experienced with the conventional electrode. A 28% decrease in discomfort with the use of the high impedance electrode was also reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.