Abstract
The PVF Tedlar is widely used for gas collection in clinical diagnostics and environmental research. However, sample collection is frequently associated with the degradation, adsorption, or transformation of sensitive chemicals. Here, we explore to what extent the Tedlar bag collection effects the composition of expired breath samples. Collected breath samples were analyzed using the EESI-MS technique after the storage time of 30 min, 1 h, 2 h, 3 h, 4 h, 5 h, and 6 h, respectively. Our results demonstrated the gradual MS signal decay after 3 h storage. The decay rate of 3 h is about 45% and 6 h is about 88%. Therefore, the Tedlar bag is suggested as a reliable breath holder on the time scale of <3 h.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Evidence-Based Complementary and Alternative Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.