The Nix Pro Colour Sensor™ (NIX) can be potentially used to measure meat colour, but procedural guidelines that assure measurement reproducibility and repeatability (precision) must first be established. Technical replicate number (r) will minimise response variation, measureable as standard error of predicted mean (SEM), and contribute to improved precision. Consequently, we aimed to explore the effects of r on NIX precision when measuring aged beef colour (colorimetrics; L*, a*, b*, hue and chroma values). Each colorimetric SEM declined with increasing r to indicate improved precision and followed a diminishing rate of improvement that allowed us to recommend r=7 for meat colour studies using the NIX. This definition was based on practical limitations and a* variability, as additional r would be required if other colorimetrics or advanced levels of precision are necessary. Beef ageing and display period, holding temperature, loin and sampled portion were also found to contribute to colorimetric variation, but were incorporated within our definition of r.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.