Abstract
A number of studies on the effects of source to image distance (SID) on dose to the patient have been published. These generally assume that the X-ray beam is collimated to a rectangular region of fixed size at the entrance surface of the patient at each SID. This is poor radiographic practice. In this work, effective doses have been determined using the commercially available Monte Carlo simulation package PCXMC. Three collimation cases were considered, namely to regions of clinical interest at the entrance and at the centre of the patient, and a fixed beam size at the image receptor. For cases of collimation to a central region of interest or to the image receptor, the saving in effective dose at increased SIDs is modest.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.