Abstract

Edge-loading of a ceramic-on-ceramic total hip replacement can lead to reproducible squeaking and revision. A patient's functional acetabular cup orientation, driven by their pelvic tilt, has been shown to be a significant factor in squeaking during hip flexion. The aim of this study was to investigate the effect of seated pelvic tilt on the contact mechanics at the ceramic bearing surface. A finite element model of a ceramic-on-ceramic total hip replacement was created. The cup was orientated at 40° inclination and 15° anteversion relative to the anterior pelvic plane. The stem was flexed 90° to replicate sitting in a chair. The model was loaded using data from in vivo measurements taken during a sit-to-stand activity. The pelvis was modelled in seven different sagittal positions, ranging from -30° to 30° of pelvic tilt, where a positive value denotes anterior pelvic tilt. Three different head sizes were investigated: 32, 36 and 40 mm. The maximum contact pressure and contact patch to rim distance were determined for each of the 21 simulations. Edge-loading (contact patch to rim distance < 0 mm) occurred with all head sizes when seated pelvic tilt was ≥10° and induced a large increase in contact pressure on the liner, with a maximum pressure exceeding 500 MPa. Edge-loading initiated at seated pelvic tilts of 7°, 9° and 5° for the 32, 36 and 40 mm heads, respectively. Patients with anterior pelvic tilts in the seated position are susceptible to posterior edge-loading. As the position of the pelvis when seated is patient specific, cup orientation should be adjusted on an individual basis to minimise edge-loading.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call