Abstract

In this article, a 3-D finite element (FE) model of human pelvic with a modular hemi-pelvic prosthesis was constructed to study the effect of screw fixation type on the biomechanics of the prosthesis. The results showed that the elimination of the screw far away from the pelvic arcuate line did not induce the instability and stress increase in the prosthesis. On the contrary, some stress in the sustain and acetabular parts decreased by 26.4% and 11.4%, respectively. In conclusion, the optimization of screw fixation can maintain the prosthesis stability and reduced stress concentration on some prosthesis parts. It was deduced that the optimization of the prosthesis could help surgeon reconstruct the pelvic joint function better and diminish the clinical time and cost.Implications for RehabilitationModular hemi-pelvic protheses are effective and popular treatments for pelvic type II resection.The elimination of screw far away from the pelvic arcuate line would induce stress descrease in some prothesis components and did not increase the instability of pelvic joint.This new prothesis could be optimized by removing the screw far away from the pelvic arcuate line and futher contribute to reduce the clinical time and cost.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.