Abstract

The effect of sample vial type and sample composition on the Ĉerenkov count rate detected from 32P and 36Cl was studied using a liquid scintillation counter. When counting was done in the noncoincident mode, glass vials allowed higher counting efficiency than plastic vials. In the coincident mode light scattering caused by polyethylene and polyproplyene vials allowed higher counting efficiency than glass vials. Highest coincident counting efficiency was from plastic minivials in a glass carrier vial. Increased solute concentration in samples caused increased counting efficiency due to changes in the refractive index of the solution. This can cause significant counting efficiency changes with no sample channel ratio change in density gradient fractions. The use of wavelength shifters is shown to be inappropriate when the sample pH varies, as this can change the fluorescent properties of the shifters and thereby the observed count rate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.