Abstract
Rare earth intensifying screen material (Gd2O2S:Tb) was added to the standard Al filtration of an oral panoramic x-ray unit, resulting in a beam capable of achieving reductions in patient dose without a loss of image quality. The added rare earth filtration technique resulted in patient dose reductions of 21-56%, depending on anatomic sites, when compared to the conventional Al filtration technique. Films generated from both techniques were measured densitometrically and evaluated by a panel of practicing clinicians. Diagnostically significant differences were minimal. The results indicate that use of rare earth filters in oral panoramic radiography is an effective means of reducing exposures of dental patients to ionizing radiation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.