Abstract

ObjectiveTo study image quality of MRI scanner using the American College of Radiology (ACR) phantom. Material and methodsImage quality of 1.5T MRI scanner was tested using ACR phantom. A standard head coil with standard restraints was used to fix the phantom head position. The phantom included seven modules for measuring MRI scanner image quality. MRI images for each module were analyzed. ResultsFor the Geometric accuracy test the inside length of the phantom was 146mm and the inside diameter was 189.4mm. For the High-contrast spatial resolution, individual small bright spots on the image were three pairs of hole arrays and were distinguishable. For the slice thickness test, the top signal ramp length was 54.4mm and bottom signal ramp length was 54.8mm so the slice thickness was 5.46mm. For the slice position accuracy, the bar length differences of intensity uniformity a value, called percent integral uniformity (PIU), was 96.15%. For the percent-signal ghosting, ghosting ratio was 0.0002. For the low contrast object detectability, the sum of numbers of complete spokes scored was 30 spokes. ConclusionsImage quality tests were very important in acceptance of any MRI scanner after installation and during maintenance. Using ACR phantom, these tests approve that the image parameters are acceptable.