Abstract

Endoscopic cyclophotocoagulation is a laser modality used to treat refractory glaucoma through the ablation of ciliary epithelium on the ciliary processes, which results in reduced aqueous production. The purpose of this study was to investigate the effect of various viscoelastics and optimal probe positioning on endoscopic cyclophotocoagulation treatment. Six different viscoelastic materials and a balanced salt solution, which was used as a control, were evaluated by placing 2 mL of each in a separate 2 mL glass tube. Laser power was measured at six distances to represent clinical placement of the laser probe in relation to the ciliary processes. The amount of energy measured with the probe at the bottom of the tube and 1 and 2 mm away was not statistically significantly different among the viscoelastic materials or the balanced salt solution. There was a gradual decrease in transmission of laser energy in all viscoelastics as the laser probe was moved farther away. To preserve the intended laser energy setting, the optimal distance between the treated tissue and the laser probe is 2 mm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.