Abstract

A prosthetic knee joint typically comprises a cobalt–chromium femoral component that articulates with a polyethylene tibial insert. A locking screw may be used to prevent micromotion and dislodgement of the tibial insert from the tibial tray. Screw loosening and back-out have been reported, but the mechanism that causes screw loosening is currently not well understood. In this paper, we experimentally evaluate the effect of polyethylene creep on the preload of the locking screw. We find that the preload decreases significantly as a result of polyethylene creep, which reduces the torque required to loosen the locking screw. The torque applied to the tibial insert due to internal/external rotation within the knee joint during gait could thus drive locking screw loosening and back-out. The results are very similar for different types of polyethylene.