Abstract

One aim of custom acetabular hip replacement sockets is to achieve fixation through a uniform cement layer of selected thickness. In vitro experiments demonstrate that curing temperatures are determined by cement thickness and position relative to the socket rim. A maximum thickness of 7 mm is indicated by interpretation of curing temperature-time relationships in terms of predicted bone necrosis. It is concluded that the results contribute to the establishment of an optimum cement layer thickness, but other factors require investigation to complement this work to gain further understanding of the problem.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.