Abstract

IntroductionTo investigate the effect of applying negative pressure (vacuum) to the periocular space on intraocular pressure (IOP) and retrobulbar pressure (RBP) by use of the Multi-Pressure Dial (MPD) system (Equinox Ophthalmic, Inc.).MethodsTwo eyes of two full body cadavers were studied. In each subject, the retrobulbar space, posterior segment and intra-goggle space were cannulated to provide direct IOP, RBP and intra-goggle pressure measurements via a pressure transducer data acquisition system. The goggles of the MPD system were placed over the eyes of each subject, and multiple test runs were performed, with negative pressure settings programmed to 5, 10 and 20 mmHg. IOP and RBP measurements were continuously obtained during each run and plotted against time for analysis.ResultsFor both subjects, the mean reduction (± standard deviation) in IOP was 1.6 ± 0.9 (10%), 3.5 ± 1.8 (23%) and 5.6 ± 2.0 (37%) mmHg at programmed negative pressure levels of − 5, − 10 and − 20 mmHg, respectively. The overall mean change in RBP (mmHg) during negative pressure application was 0.02 ± 0.14 at − 5 mmHg, 0.03 ± 0.19 at − 10 mmHg and − 0.01 ± 0.18 at − 20 mmHg. In both subjects, the magnitude of RBP change during application of negative pressure fell below the uncertainty of the measurement system.ConclusionsThe application of negative pressure to the periocular space with the MPD decreases IOP but does not affect RBP.Supplementary InformationThe online version contains supplementary material available at 10.1007/s40123-021-00442-2.