Abstract

Albumin-scavenging surfaces decorated with n-alkyl chains represent an established strategy for blood-contacting applications. To evaluate this concept, a set of poly (2-hydroxyethyl methacrylate) (pHEMA) films modified with different amounts of octadecyl isocyanate (C18) was investigated in an in vitro hemocompatibility assay using freshly drawn human whole blood. In addition, the hydrogel materials were thoroughly characterized with respect to changes in wettability and elasticity, which accompanied the gradual chemical modification of pHEMA. An increase of the surface C18 content induced enhanced hydrophobicity and stiffness. Immobilization of C18 chains was found to substantially reduce the coagulation activation and the complement activation by the pHEMA films. Platelet adhesion and degranulation (PF4 release) were similar on the modified and the unmodified pHEMA. Platelet adhesion to pHEMA hydrogels was lower than the polytetrafluoroethylene reference. We conclude that the immobilization of octadecyl chains improved the hemocompatibility of pHEMA materials under conditions that might be encountered in low shear blood flow.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.