Abstract

The purpose of this study was to show a change in proximal femur surface strains following total hip arthroplasty and after the addition of BoneSource hydroxyapatite bone cement in the proximal region of an instrumented femur and to measure the surface strain on the proximal body. Seven third-generation composite femurs (Pacific Research Laboratories, Vashon, Wash) were instrumented with 12 uniaxial strain gages, 6 gages on the anterior face, and 6 gages on the posterior face of each femur. All femurs exhibited stress shielding since the strains in the proximal region were drastically reduced. There was a large decrease in strain in the mid-shaft region and small changes in strain in the distal region. The surface strains on the modular implant were relatively low.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.