Abstract
The performance of a Bayard–Alpert type ionization gauge under the influence of weak magnetic fields (<60 G) has been studied in detail. The sensitivity of this BA gauge varied with (1) the magnitude of the magnetic field, and (2) the direction of the magnetic field, especially the angle between the direction of the emitting electrons and the direction of the magnetic field. If a BA gauge is required to operate in a magnetic field, the optimum condition could be obtained by carefully choosing the mounting orientation relative to the direction of the magnetic field.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.