Abstract

This study was performed to determine how the use of an introducer affects the extent to which a needle deflects during a spinal or combined spinal-epidural injection. A polystyrene block was used to simulate the paraspinal area of the back. A line was drawn perpendicular to the edge of the block to use as a guide and to measure the deflection. The use of an introducer needle decreased the deflection in all the bevelled needles (p < 0.001). Depending on the direction of both the bevels, the deflection decreased as the introducer bevel was changed from the same direction, to right-angles to bevel direction and then to a direction opposite to that of the spinal needle (p < 0.05). Deflection was decreased when a thick introducer was used (p < 0.001). The use of an introducer increased the deflection of the pencil-point needle only in the deflection direction of the introducer (p < 0.001). The 18-gauge Tuohy needle with a "backhole" deflected more than the corresponding needle without a backhole (p < 0.001), and the spinal needle inserted through the Tuohy needle with a backhole deflected more (p = 0.002). Besides the tip type and gauge, the deflection of a spinal needle depends upon the use of introducer, its gauge and bevel direction. The deflection of a Tuohy needle depends upon its design, gauge and the presence of a backhole.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.